|
Microscopy resources, techniques, and materials for classroom observations:
 |
This is a compound microscope equipped with an adjustable substage condenser
lens and an illumination field-limiting aperture (just above the light source).
|
Microscopes
One or more compound microscopes, preferably binocular, with 10
X ocular lenses, and 6, 10, and 40 X objective lenses will be needed. These
will give a total magnification of 60, 100, and 400X, respectively. Students
can be asked to figure out magnification by multiplying the magnification factors
for ocular and objective lenses. Microscopes that have adjustable field limiting
apertures and condensers are best, because these can be aligned for optical
image viewing. Taking pictures of
plankton is enjoyable, so it is nice to also
have a photographic system, especially a digital camera, video camera, or digital
video camera attached to a microscope.
Aligning a compound microscope
Better images will be observed if the microscope is aligned and illumination
field adjusted for each magnification. To accomplish this, place a prepared
elide of a thin object, such as
diatoms or sectioned plant or animal tissue
on the microscope stage and focus the image at the desired magnification. Then
close the illumination field limiting aperture at the bottom of the microscope
until you can see its edges through the lenses.
 |
This image of the field limiting aperture needs
to be focused, by moving the condenser height, and it needs to be centered
by moving the image laterally. |
Now move the condenser knob until the edges of the aperture are sharp; you
may see a change from red to blue flare at the edges. Set the condenser at this
sharp, flare point. Now move the two knobs that align the condenser until the
image of the field limiting aperture is centered
This field limiting aperture is correctly positioned. |
 |
Finally, open the field limiting aperture until the edges are just out of view.
This setting should be adjusted each time a different magnification is used.
|